28000信号调理系统
高精准28000 信号调理系统单机可提供高达256 通道的可编程传感器信号调理。它包含桥路、电荷、IEPE w/TEDS、 电压(滤波放大器)、应变、热电偶、RTD/电位计、频率信号或者其他传感器的在内的信号调理。
8-slot and 16-slot rugged and transportable mainframes feature integrated cooling and field swappable power supplies. 
System Benefits
-
Manage hundreds of channels with a mix of sensors
-
Friendly system setup via local GUI or Ethernet network interface
-
Assure system integrity before performing measurements Reduce life-cycle costs
-
Upgrade and expand as equipment and requirements change Count on unsurpassed performance and reliability
Features
-
Mix up to 256 channels of fully programmable transducers in a single system
-
Graphical User Interface (GUI) for system control
-
Intelligent gain and system scaling algorithms
-
Test input and output monitor busses
-
Go/no-go test with diagnostics to be used before tests
-
Rigorous factory acceptance test for maintenance
-
Field swappable power supplies
-
Built-in temperature and power supply monitoring with alarms
信号调理类型
-
桥路输入
-
动态应变
-
RTD/电位计
-
热电偶
-
频率电压转换
-
电荷
-
电压
-
转动机械
-
IEPE w/TEDS 传感器
Filter Characteristics
The 28000 System has a variety of high-performance filter characteristics available for HP, LP or BP Precision filtering, as well as, traditional filter types such as Butterworth and Bessel characteristics and others.

The following card selection guide is provided to aid in identifying the 28000 Signal Conditioner cards that meet your requirements.
|
Model |
28104A |
28208 |
28302B |
27304 |
28314 |
28316B |
28454A |
28524 |
28608B |
|
Type |
Bridge |
Thermo- couple |
SE or Diff Charge, IEPE or Velocity Amplifier |
Charge/ Voltage or IEPE Amplifier |
Dynamic Charge or IEPE w/LDTEDS |
IEPE w/LDTEDS |
Dynamic or Static Strain, RTD |
F-to-V, Pulse Rate |
Prog. Filter/ Amplifier |
|
Channels/ Card |
4 |
8 |
2 |
4 |
4 |
16 |
4 |
4 |
8 |
|
Channels/ System |
64 |
128 |
32 |
64 |
64 |
256 |
64 |
64 |
128 |
|
Application Transducer |
Strain, load cell, pot., pressure, low level AC, DC or any bridge |
Temper- ature, low-drift, low level DC or AC amp/filter |
Rotating machinery vibration, G, ips and mil outputs |
Charge or IEPE Accels, vibration |
Dynamic measure-ments for shock and vibration |
IEPE Accels w/LDTEDS, AC filter/ amp |
Dynamic or static strain with 2-wires, RTD, bridge, low-level AC or DC |
Frequency counter, pulse rate, flow rate, Hall effect sensors |
Low-level AC or DC amps and LP, HP or BP filtering |
|
Interface |
2-10- wire; 1, 2, 4 arm |
3-wire Iso- thermal Block |
Lemo or Coaxial |
Coaxial |
Coaxial |
2-wire w/shield |
2/4-wire |
2-wire w/shield |
2-wire w/shield |
|
Excitation |
Balanced constant voltage 0-20.475 VDC (30 mA) |
N/A |
Constant current (IEPE), 8.5 mA |
Constant current (IEPE), 1-12.75 mA |
Constant current (IEPE), 1-25 mA |
Constant current (IEPE), 0, 1, 2, 4, or 8 mA |
Balanced diff constant current 0-25.5 mA |
N/A |
N/A |
|
Vcm |
10 V |
10 V |
10 V |
3.2 V |
10 V |
N/A |
10 V |
10 V |
10 V |
|
Model |
28104A |
28208 |
28302B |
27304 |
28314 |
28316B |
28454A |
28524 |
28608B |
|
Bandwidth |
700 kHz |
1 kHz |
5 Hz to 20 kHz |
200 kHz |
500 kHz |
100 kHz |
700 kHz |
1 Hz to 50 kHz |
500 kHz |
|
Gain |
x1/16 to x8192 |
x1, 10, 100, 1000 |
Sensor sensitivity: 0.1 pC/g to 110 pC/g |
x1/16 to x4096
|
x1/16 to x8192 |
x0.5 to x128 |
x1/16 to x8192 |
Input range 10 mV to 100 V |
x1/4 to x8192 |
|
Filter Poles |
4, 8 |
3 |
12 |
4, 6, 8 |
4, 8 |
4 |
4, 8 |
4 |
4, 8 |
|
Filter Types |
Pulse/Flat LP |
BE |
LP, HP, Band-Pass |
BE, BU, EL, LP, TD |
Pulse/Flat LP |
BE, BU, EL, TD |
Pulse/Flat LP |
Band-Pass |
Pulse/Flat LP, HP, BP |
|
Fc Range |
Flat Mode: 2 Hz to 204.6 kHz
Pulse Mode: 1 Hz to 102.3 kHz |
1, 10, 100 Hz, Prog. |
HP: 5 Hz to 1,275 Hz
LP: 50 Hz to 12,750 Hz |
Available in many ranges:
2-Range models; 1 Hz to 127.5 kHz
Single 255:1 range
Fixed range models |
Flat Mode: 2 Hz to 204.6 kHz
Pulse Mode: 1 Hz to 102.3 kHz |
100 Hz to 30 KHz, Prog. |
Flat Mode: 2 Hz to 204.6 kHz
Pulse Mode: 1 Hz to 102.3 kHz |
1 Hz to 50 KHz, Prog. |
Flat Mode: 2 Hz to 204.6 kHz
Pulse Mode: 1 Hz to 102.3 kHz |
|